OG05 5m3 medical oxygen gas generation plant

Group medical oxygen generator
PriceUS $12000 / piece
Min. Order 1 piece
Terms of Payment L/C, T/T
Update Time 2016-09-17
Item specifics
Oxygen flow capacity5,0Nm3/h
Oxygen purity95%
Air inlet pressure7.5bar
Oxygen outlet pressure6.1+/-0.5bar
Air consumption56,4Nm3/h
Power supply230V-1phase-50HZ
Dimensions1048x560x1672mm
Weight458kg
  • Description
  • Product review

OG05-5m3/h,95% purity medical oxygen gas generation plant

DESCRIPTION

The OG series oxygen generators extract theavailable oxygen in the ambient air from theother gases by applying the Pressure Swing Adsorption (PSA) technology. During the PSA process compressed, cleaned ambient air is led toa molecular sieve bed, which allows the oxygen topass through as a product gas, but adsorbs othergases.

The sieve releases the adsorbed gases to theatmosphere, when the outlet valve is closed andthe bed pressure returns to ambient pressure.Subsequently the bed will be purged with oxygenbefore fresh compressed air will enter for a newproduction cycle.

In order to guarantee a constant product flow,oxygen generators use modules of two molecularsieve beds, which alternatively switch betweenthe adsorption and the regeneration phase.Under normal operating conditions and withcorrect maintenance the molecular sieve bedswill have an almost indefinite lifetime.


STANDARD EQUIPMENT

• Feed Air Filters

•Adsorber Vessel Module(s)

• Pneumatic Valves with SS316L Bodies

• Internal Piping & Fittings in SS316

• Exhaust Mufflers

• Oxygen Pressure Regulation

• Local Instrumentation

• Control System with Allen-Bradley PLC

• Pressure Switch for Automated Idle-Mode


OPTIONAL EQUIPMENT

• Dual Bank Unit (max. 3)

• Oxygen Analyzer

• Electronic Product Flow Meter

• Touch screen or Semi-Graphical Operator Interface

• Telemetry, even through GSM or Ethernet

• Sterile Filters

• External O2 Analyzing System incl. PDP, CO/CO2

• MedOx External Control System for Duplex Units

• Oxygen Booster with Cylinder Filling System


PERFORMANCE

Type

Inlet pressure

Discharge pressure

OXYGEN PURITY [%]

[barg]

[barg]

90

93(1)

95

OG03

7,5

6,1 (± 0,5 bar)

2,6

2,5

2,5

Feed air consumption [Nm3/h]

28,2

28,2

28,2

OG05

7,5

6,1 (± 0,5 bar)

5,3

5,1

5,0

Feed air consumption [Nm3/h]

56,4

56,4

56,4

OG08

7,5

6,1 (± 0,5 bar)

8,0

7,8

7,6

Feed air consumption [Nm3/h]

84

84

84

OG10

7,5

6,1 (± 0,5 bar)

10,5

10,2

10,0

Feed air consumption [Nm3/h]

111,6

111,6

111,6

OG13

7,5

6,1 (± 0,5 bar)

13,1

12,7

12,5

Feed air consumption [Nm3/h]

139,2

139,2

139,2

OG15

7,5

6,1 (± 0,5 bar)

15,5

15,1

14,8

Feed air consumption [Nm3/h]

168

168

168


TECHNICAL DATA

Type

Connection

Dimensions [mm]

Weight

In

Out

L

B

H(H)*

kg

OG03

1''

1''

780

560

1672 (1510)

306

OG05

1''

1''

1048

560

1672 (1510)

458

OG08

1''

1''

1316

560

1672 (1510)

609

OG10

1''

1''

1584

560

1672 (1510)

759

OG13

1''

1''

1852

560

1672 (1510)

910

OG15

1''

1''

2120

560

1672 (1510)

1061

(1) Purity according to the Oxygen 93 Monograph of European Pharmacopoeia 7.1 and USP23 and conform ISO10083 standard.

Flow rates at standard atmospheric conditions (20 °C/70 °F,1013 mbar/14,7psi and 60% RH).


    No Review